QIAcuity Probe PCR Kit

供与 QIAcuity 数字 PCR 仪器一起使用

S_9796_dPCR_QIAcuity_Probe_PCR_Kit

✓ 全天候自动处理在线订单

✓ 博学专业的产品和技术支持

✓ 快速可靠的(再)订购

QIAcuity Probe PCR Kit (1 ml) icon_0368_ls_gen_eco_friendly-s

目录编号 / ID.   250101

1 ml 4x 浓缩 QIAcuity Probe Mastermix,2 x 1.9 ml 水
体积
1 ml
25 ml
QIAcuity Probe PCR Kit 旨在用于分子生物学应用。该产品不能用于疾病诊断、预防和治疗。

✓ 全天候自动处理在线订单

✓ 博学专业的产品和技术支持

✓ 快速可靠的(再)订购

特点

  • 适用于单重和最多 5 重反应
  • 4x 浓缩预混液,可加载更多样本
  • 针对在 QIAcuity Nanoplate 中的微流体使用优化
  • 更少的非特异性 PCR 扩增
  • 符合 REACH 法规

产品详情

QIAcuity Probe PCR Kit 包含一份用在 QIAcuity Nanoplate 微流体中优化的 4x 浓缩即用型预混液。该试剂盒提高了基于探针法的数字 PCR 的特异性和效率,可提供准确的单重或最多 5 重的定量分析结果。QIAcuity 探针法预混液经过优化,提高特异性,从而可在 QIAcuity dPCR 仪器上准确定量 gDNA 或 cDNA。

该试剂盒应与 QIAcuity Digital PCR SystemQIAcuity Nanoplate结合使用。

您想要进一步了解该产品并让我们的 dPCR 专家联系您吗?请在这里登记,我们将很快与您联系。

绩效

卓越的性能
基于水解探针法的 dPCR, QIAcuity 预混液中使用了最新版的 QIAGEN 高品质 DNA 聚合酶。QIAGEN 的专有技术与经过良好验证的专门针对纳米微孔板微流体优化的缓冲液技术及新的 QuantiNova DNA 聚合酶的独特结合,可在灵敏度、重复性及效率等方面获取高度一致的结果。

使用 QIAcuity Probe PCR Kit 可进行基于探针法的单重到五重 dPCR检测
QIAcuity Probe PCR Kit 中的独特预混液可让您在 QIAcuity Nanoplate 单孔中对最多 5 个丰度大为不同的靶标基因进行准确定量。这可节省时间和金钱,降低所需的样本材料用量。此外,双重或多重 PCR获取的数据与单重 PCR 中获得的数据具有一致性。

反应稳定性可长达 100 个小时
QIAcuity PCR 混合物可在 30°C 下存放长达 100 个小时,而不会影响后续反应的性能。优异的稳定性——即便在室温不使用冷却剂的情况下,长时间存放后仍保持性能的稳定性,使得 QIAcuity Probe PCR Kit 成为高通量反应体系构建及孔板堆叠处理时的理想选择。

原理

QIAcuity Probe PCR Kit 采用新型抗体介导的热启动机制,可确保 cDNA 或 gDNA 单重或多重分析时的高特异性。在低温下,QuantiNova DNA 聚合酶在 QuantiNova 抗体和 QuantiNova Guard(一款具有复合物稳定化作用的新型添加剂)的作用下保持在失活状态。这提高了热启动的严格性并防止非特异性退火引物的延伸及引物二聚体的形成。在温度升高到 95°C 的两分钟内,QuantiNova 抗体和 QuantiNova Guard 将发生变性,QuantiNova DNA 聚合酶将被激活,从而启动 PCR 扩增。

有关纳米微孔板中 dPCR 反应的原理说明,请参见此处

程序

与 qPCR 实验一样,样本制备包括将预混液、探针和引物转移到一个 96 孔或 24 孔纳米微孔板中,随后添加样本。该系统将微分化、热循环和成像集成到单一全自动化的仪器上,使用户可在 2 小时内获得结果。您可在 Suite Software 上执行分析,该软件可针对您的靶序列及质量控制(例如阳性样本或无模板对照 [NTC])给出以每微升拷贝数表示的浓度。该分析还可在同一局域网 (LAN) 上的远程计算机上执行。

应用

QIAcuity Probe PCR Kit 与 QIAcuity Digital PCR SystemQIAcuity Nanoplate 的结合使用可让您在各种应用中进行 cDNA 靶标和 gDNA 定量分析,这些应用包括:

  • 罕见突变检测
  • 拷贝数变异分析
  • 基因表达分析
  • 病原体检测
  • 基因分型
  • miRNA 研究
  • 细胞和基因治疗
  • 宿主细胞 DNA 残留定量 
  • 废水监测

辅助数据和图表

常见问答

Is it possible to change voltage set-up from 110V to 230V on the QIAcuity instruments?

This is not needed. The QIAcuity is equipped with a flexible power supply technology and operates within a range of 100–240V AC, 50/60 Hz, 1500 VA (max). 

FAQ-3761
Can I see error codes on the instrument touchscreen?

The instrument software GUI shows error codes including a description and information how to resolve the error. The instrument touchscreen shows an alarm icon in the upper right corner that turns red in case of an instrument failure. Accessing the System Status in the Tool tab allows users to clear errors. Rebooting of the instrument is required to complete the removal of the error.  Please do not skip this step. You may always contact QIAGEN Technical Services in case of any question. 

FAQ-3763
What is the scope of a regular maintenance of the QIAcuity?

The user manual contains instructions on how to perform a regular cleaning and decontamination, and how to replace air filters on the QIAcuity instruments. A regular maintenance reduces the dust in the instrument and therefore minimizes the presence of dust particles on the nanoplate, which might interfere with the plate analysis.

FAQ-3765
What is the impact of not applying the latest VPF? Can I reanalyze previously obtained results after installing the latest VPF?

If you had run a nanoplate for which the installed VPF misses the specific factor, the software will notify you. If you then analyze without the specific VPF, the impact depends on the variation of the partition volume of the new Nanoplate batch compared to the latest. Typically this variation is ±6–7% (approx. 5% CV over the entire plate). The analysis may be repeated after updating the VPF file. After installing the latest VPF and re-analysis of the run, a copy of the plate is generated in the QIAcuity Software Suite including the new results. 

FAQ-3769
Can I see on the QIAcuity Software Suite report file if the VPF (Volume Precision Factor) has been used or not?

Yes, the report includes a notification if the matching VPF was missing and, therefore, not applied to the analysis. If the matching VPF was applied there is no notification on the report. 

FAQ-3770
Is it necessary to reanalyze a plate with VPF (Volume Precision Factor) that was already processed using a QIAcuity instrument that was purchased in 2020?

If you had analyzed your nanoplates without VFP, the impact depends on the variation of the partition volume of the new nanoplate batch compared to the latest. The VPF reduces the CV from approximately 5% to 2%. We recommend to reanalyze results in case the data originated from different wells (e.g., copy number variations or gene expression data sets for which the reference sample was measured in a different well). Results obtained across different plates should also be r-analyzed. A reanalysis is not required for assay data that were analyzed within the same well (e.g., mutation rate determination using two channels within the same well).

FAQ-3771
Can I prepare a dPCR reaction directly in QIAcuity Nanoplate?

A standard PCR plate is required to set up dPCR reaction before transferring it to the nanoplate to ensure a proper mixing of the reaction mix before partitioning. 

FAQ-3774
Can I use a custom master mix instead of a QIAGEN master mix?

QIAGEN master mixes are optimized for nanoplate microfluidics and are recommended to be used with our dPCR system. They also include an optimized reference dye required for proper analysis.

FAQ-3777
How to prepare DNA prior to dPCR?

All DNA samples used in reaction mixes should show similar quality and quantity, which can easily be assessed using UV spectrophotometry. DNA samples with an average length of ≥20 kb (e.g., genomic DNA purified via spin column with silica membrane) should be fragmented by restriction digestion before partitioning. Enzymatic fragmentation of larger DNA ensures an even distribution of template throughout the QIAcuity Nanoplate, which in turn leads to an accurate and precise quantification.

FAQ-3778
What are the storage conditions and expiry date of QIAcuity consumables?

The QIAcuity Probe PCR Kit should be stored immediately upon receipt at –30 to –15°C in a constant-temperature freezer and protected from light. The QIAcuity Probe PCR master mix can also be stored protected from light at 2–8°C. Components are stable for 12 months, unless otherwise indicated on the label. 
The QIAcuity EG PCR Kit should be stored immediately upon receipt at –30 to –15°C in a constant-temperature freezer and protected from light. The QIAcuity EG PCR master mix can also be stored protected from light at 2–8°C. Components are stable for 6 months, unless otherwise indicated on the label. 
The QIAcuity Nanoplates does not have expiry date and are stable for at least 1 year when stored at RT. 

FAQ-3780
Can I use the QIAcuity Nanoplate in more than one runs?

The plate is designed for a single use run. For example, even if only 30 samples are loaded into the 96-well plate, a whole plate will be sealed by the roller. It can't be unsealed and used for another run. The QIAcuity Software won’t allow to set up a separate experiment for the same nanoplate to avoid that previously processed plates being not partitioned a second time.

FAQ-3781
Can I optimize a dPCR assay on the QIAcuity using gradient temperature?

An essential temperature gradient functionality was introduced with software version 2.5. When updating older software versions to 2.5, each QIAcuity instrument will offer the temperature gradient and may be used to find the best cycling temperature for new dPCR assays. When running a QIAcuity Four or a QIAcuity Eight, all plates may have their own temperature profile, including the option for a temperature gradient.

FAQ-3783
What is the VPF (Volume Precision Factor)?

The VPF provides a set of well-specific and molding form-specific factors used to specify the exact reaction volume of a nanoplate, thus increasing the concentration calculation of each well. 

FAQ-3784
When and how often do I need a new VPF (Volume Precision Factor)?

 In general, nanoplates provide partitions of fixed sizes that enable a very precise way of sample concentration calculation. If a new molding form is used for nanoplate manufacturing, potential variation of partition sizes can be addressed by applying the molding form-specific VPF. Thus, each time a new molding form is used, a new VPF is created and made available. Currently, the VPF is updated once every 3–6 months.

FAQ-3785
Is a standard curve needed in dPCR?

In dPCR we measure the absolute concentration of targets at endpoint reaction. Concentrations of unknowns can be determined based on dPCR results observed (number of negatives, number of positives, and partition volume analyzed).

FAQ-3786