What sample types should be processed with the low content (LC) and high content (HC) protocols for DNA extraction on the QIAsymphony SP?
Tissues will give increased DNA yields when processed with the high content protocol, but the low content protocol in combination with a small elution volume (50µl) may be used if high DNA concentration is required. For FFPE tissue we recommend using the low content protocol.