RNase-Free DNase Set

For DNase digestion during RNA purification

S_1299_GEF_LE0131
Need bulk, customized or optimized products for commercial purposes? We also offer support with logistics, compliance and more. Reach out to cooperate with QIAGEN Strategic Partnerships & OEM

RNase-Free DNase Set (50)

Cat. No. / ID:   79254

1500 Kunitz units RNase-free DNase I, RNase-free Buffer RDD, and RNase-free water for 50 RNA minipreps
270,00 $SG
Log in To see your account pricing.
Preparations
50
250
The RNase-Free DNase Set is intended for molecular biology applications. This product is not intended for the diagnosis, prevention, or treatment of a disease.
Need bulk, customized or optimized products for commercial purposes? We also offer support with logistics, compliance and more. Reach out to cooperate with QIAGEN Strategic Partnerships & OEM

Features

  • Compatible with RNeasy procedures and the QIAamp RNA Blood Mini Kit
  • Efficient on-column digestion of DNA during RNA purification

Product Details

The QIAGEN RNase-Free DNase Set is guaranteed RNase-free, quality-controlled, and optimized for use with RNeasy procedures and with QIAamp RNA Blood Mini procedures. Generally, DNase digestion is not required for RNA purified with RNeasy Kits since the silica-gel–membrane, spin-column technology efficiently removes the majority of the DNA without DNase treatment. However, more complete DNA removal may be necessary for certain RNA applications that are sensitive to very small amounts of DNA. Buffer RDD, included in the set, is optimized for on-column DNase digestion of 15 minutes at 20–30°C. The buffer is also well-suited for efficient DNase digestion in solution. The RNase-Free DNase Set provides efficient on-column digestion of DNA during RNA purification from cells and tissues using RNeasy Kits and the QIAamp RNA Blood Mini Kit. The DNase is efficiently removed in subsequent wash steps. For DNase treatment when using the RNeasy 96 Kit, please contact QIAGEN Technical Services or your local distributor for a separate, optimized protocol. The QIAGEN RNase-Free DNase Set is delivered as a stable, lyophilized enzyme. The RNase-Free DNase Set provides 1500 Kunitz units.

Performance

One Kunitz unit is defined as the amount of DNase I that causes an increase in A260 of 0.001 per minute per milliliter at 25°C, pH 5.0, with highly polymerized DNA as the substrate (Kunitz, M. [1950] Crystalline desoxyribonuclease. I. Isolation and general properties, spectrophotometric method for the measurement of desoxyribonuclease activity. II. Digestion of thymus nucleic acid (desoxyribonucleic acid): the kinetics of the reaction. J. Gen. Physiol. 33, 349 and 363).

Resources

Kit Handbooks (1)
For DNase treatment with QIAGEN or PreAnalytiX RNA purification kits  
Safety Data Sheets (1)
Download Safety Data Sheets for QIAGEN product components.
Certificates of Analysis (1)

FAQ

How can I obtain DNA-free RNA using an RNeasy Midi or Maxi Kit?
The RNeasy Midi/Maxi Handbook contains a protocol for the use of the RNase-Free DNase Set for on-column DNA digestion on RNeasy midi or maxi spin columns. Incubation times and reagent volumes have been modified from the standard RNeasy Mini procedure. See Appendix E of the RNeasy Midi/Maxi Handbook for the detailed protocol.

FAQ ID -143
I accidentally stored Buffer RDD of the RNase-Free DNase Set at°C. Will it still function?
Yes, buffer RDD of the RNase-Free DNase Set will still work. Please make sure that the buffer is thawed completely without any precipitates before using it. If precipitates are visible, the buffer should be slightly heated.
-20
Can I use the RNeasy MinElute Cleanup Kit to clean up my in vitro transcription reaction?
Yes. RNeasy MinElute Cleanup Kit can also be used to clean up RNA following DNase digestion, or from crude RNA preparations following organic extraction or alcohol-precipitation.
FAQ ID -429
What is the composition of Buffer RDD?
The exact composition of Buffer RDD is proprietary. Buffer RDD is an important component of the RNase-Free DNase Set, which is used in combination with most RNeasy Kits. The composition and salt concentration of Buffer RDD provides efficient on-column digestion of DNA and also ensures that the RNA remains bound to the column.
FAQ ID -2800
Can the RNase-Free DNase Set be used for DNase digestions of RNA in solution?

Yes. Even though buffer RDD in the RNase-Free DNase Set is optimized for on-column DNase digestion, the buffer is also well-suited for efficient DNase digestion in solution. Please note that the reaction must be cleaned up after the off-column DNase digest to remove the enzyme and buffer RDD, which will interfere with subsequent RT reactions.

A protocol for in-solution DNase digestion using the RNase-Free DNase Set can be found in Appendix C of the RNeasy MinElute Cleanup Handbook. For subsequent RNA Cleanup, use either the RNeasy MinElute Cleanup Kit, or follow the instructions for RNA Cleanup in the RNeasy Mini Handbook.

 

FAQ ID -619
How should I reconstitute the lyophilized RNase-Free DNase Set with the RNeasy 96 kit?

RNeasy 96 Kit procedure requires two RNase-Fee DNase sets (cat. no. 79254) for each 96-well plate.

 

1. To obtain a concentration of 1.8 Kunitz/µl, dissolve each vial of lyophilized DNase in 833 µl of RNase-free water and combine the reconstituted enzyme solution from both vials into one vial.

2. Take 1 ml of this DNase I solution and mix with 7 ml of Buffer RDD* to prepare a DNase I incubation mix immediately before starting the RNeasy 96 protocol. Vortex briefly and keep on ice until use.

3. Use 80 µl of the DNase I incubation mix (from step 2 above) for each well of the RNeasy 96 plate, to perform on-column DNase digestion, according to instructions in the RNeasy 96 Handbook.

 

*The RDD Buffer Set for RNeasy 96 is available upon request.  Please contact QIAGEN Technical Service for information on ordering.  Buffer RDD from QIAGEN is optimized for DNase I digestion on the RNeasy membrane.

FAQ ID -3003
How can I avoid or remove genomic DNA contamination from the total RNA preparation?

Carry out all procedures in a "DNA-free" workspace (see FAQ 2654). Be sure to include any DNase treatment steps in the recommended RNA isolation procedure or treat RNA separately with RNase-free DNase followed by repurification using a spin-column based method. RNeasy Mini Kit can be easily combined with RNase-free DNase. Alternatively, kits like the RNeasy Plus Universal Tissue already include a DNA removal step. Be sure to double both the units of enzyme and the incubation time recommended by the RNase-free DNase manufacturer. To minimize DNA contamination in your RNA preparations, and avoid the need for supplemental DNase treatments, we recommend using the RT2 First Strand Kit, which includes a highly efficient genomic DNA elimination step before reverse transcription.

 

NOTE: Our Chemistries are not compatible with AMBION's Turbo DNA-Free Kits.

FAQ ID -2662
How can I ensure complete genomic DNA removal when using the RNase-Free DNase Set?

To ensure efficient gDNA removal when doing an on-column digest using the RNase-Free DNase Set in combination with RNeasy Mini the following factors are crucial:

  • prevent overloading by adjusting the amount of starting material to no more than the maximum amounts recommended in the RNeasy Mini Handbook
  • ensure complete disruption and homogenization of the starting material as instructed in the section 'Disruption and homogenization of starting materials' of the handbook
  • strictly follow the protocol for on-column DNase Digestion in Appendix D of the RNeasy Mini Handbook (you can let wash buffer RW1 incubate on the column for 3-5 minutes before centrifuging to enhance removal of excess gDNA prior to applying the enzyme)

In the rare case that trace amounts of genomic DNA are still detectable in sensitive downstream applications such as e.g., realtime RT-PCR, an in-solution digest using the RNase-Free DNAase set can be performed. Instructions are presented in Appendix C of the RNeasy MinElute Cleanup Handbook.

Alternatively, a second on-column digest can be carried out in future preparations, immediately following the RW1 wash after the first incubation with DNase.

FAQ ID -1087