What possible reasons result in a sensitivity lower than that given in the artus® user handbook?
• The artus® Master was not completely thawed and as a result the concentration is incorrect.
• The artus® Master was not mixed sufficiently by vortexing.
• The Mg2+ solution was not completely mixed/thawed (where applicable) before being added to the artus® Master.
• The solutions were not sufficiently mixed and then centrifuged briefly.
• The PCR setup was not carried out on ice or in a cooler.
• The solutions were frozen and thawed too often (if possible do not thaw more than twice).
• During transportation the kits were not kept frozen permanently.
• Pipettes are not calibrated correctly.
• Pipette tips do not fit the pipettes.
• An extraction procedure other than those recommended was used.
• The extraction volume has been increased.