therascreen EGFR Plus RGQ PCR Kit

For detection and identification of 42 mutations in exons 18, 19, 20 and 21 of the EGFR gene

S_1212_6_MDx_CG_theraEGFRPlusRGQPCR_CEIVD

✓ 24/7 automatic processing of online orders

✓ Knowledgeable and professional Product & Technical Support

✓ Fast and reliable (re)-ordering

therascreen EGFR Plus RGQ PCR Kit (24)

Cat. No. / ID:   874611

For 24 reactions: T790M & L861Q Mix, Insertions & G719X Mix, L858R & C797S Mix, Deletions & S768I Mix, PCR Master Mix, EGFR Positive Control, RNase/DNase-free Water
ProductInstrumentSoftware
therascreen EGFR Plus RGQ PCR Kit (24)
QIAamp DSP DNA FFPE Tissue Kit (50)
QIAamp DSP Circulating NA Kit
QIAsymphony DSP Circulating DNA Kit (192)
Rotor-Gene Q MDx HRM System
Rotor-Gene Q MDx HRM Platform
Rotor-Gene AssayManager
The therascreen EGFR Plus RGQ PCR Kit is intended for in vitro diagnostic use.

✓ 24/7 automatic processing of online orders

✓ Knowledgeable and professional Product & Technical Support

✓ Fast and reliable (re)-ordering

Features

  • Detection of TKI resistance biomarkers, including T790M and C797S
  • Run FFPE and liquid biopsy samples in the same PCR run
  • Reliable, sensitive and specific detection of clinically relevant EGFR mutations
  • Automated sample preparation option with QIAsymphony SP
  • Automated analysis and next-day result reporting with Rotor-Gene Assay Manager software

Product Details

The therascreen EGFR Plus RGQ PCR Kit is an in vitro diagnostic real-time PCR test intended for the qualitative detection and identification of 42 mutations in exons 18, 19, 20 and 21 of the epidermal growth factor receptor (EGFR) gene in DNA samples extracted from plasma or formalin-fixed paraffin-embedded (FFPE) tumor tissue taken from non-small cell lung cancer (NSCLC) patients.

Furthermore, the kit also indicated for the semi-quantitative measurement of mutations in exons 18, 20 and 21 of the EGFR gene in DNA samples extracted from human plasma as an aid in the clinical management of NSCLC patients.

The test is intended to be used as an aid in selecting patients with NSCLC for therapy with an EGFR tyrosine kinase inhibitor (TKI).

Performance

When compared to the analytical reference method, there was a high level of concordance between therascreen EGFR Plus RGQ PCR Kit and the analytical accuracy method(s). The reference and discrepancy resolution methods used were: PNA qPCR, Sanger bi-directional sequencing; next-generation sequencing, therascreen EGFR RGQ PCR Kit V2 (cat. no. 874111) and therascreen EGFR Plasma RGQ PCR Kit (cat. no. 870311).

Results were analyzed to assess the positive percent agreement (PPA), negative percent agreement (NPA) and overall percent agreement (OPA) regarding EGFR mutation status (mutant; MT or wild-type; WT) and EGFR target (mutation identification) for FFPE and plasma samples between therascreen EGFR Plus RGQ PCR Kit and respective reference method and following discrepancy resolution method.

FFPE samples

In the study 170 FFPE samples were tested, of which 48 gave valid interpretable results (148 sample status and 155 target status).

When the results of the therascreen EGFR Plus RGQ PCR Kit were compared to the results of the respective reference method, four EGFR sample status (MT or WT) showed discordance. Following discrepancy resolution method analysis, the number of discordant samples statuses (MT or WT) decreased to one discrepant, false-negative sample status. The PPA, NPA and OPA with the corresponding two-sided 95% confidence intervals (CI) are summarized in Table 1 and Table 2.

Table 1. Analysis of agreement of overall mutation status per sample: therascreen EGFR Plus RGQ PCR Kit and reference method comparison for FFPE samples

    Lower limit 95% CI Upper limit 95% CI
OPA 97.30% 93.22% 99.26%
PPA (sensitivity) 93.65% 84.53% 98.24%
NPA (specificity) 100.00% 95.75% 100.00%

Table 2. Analysis of agreement of overall mutation status per sample after discordant investigation for FFPE samples

    Lower limit 95% CI Upper limit 95% CI
OPA 99.32% 96.29% 99.98%
PPA (sensitivity) 98.33% 91.06% 99.96%
NPA (specificity) 100.00% 95.89% 100.00%

Plasma samples

In the study, 106 plasma samples were tested and 106 gave valid interpretable results (106 sample status and 121 target status).

When the results of therascreen EGFR Plus RGQ PCR Kit were compared to the results of the respective reference method, nine EGFR sample status (MT or WT) showed discordance. Following discrepancy resolution method analysis, the number of discordant samples status (MT or WT) decreased to three discrepant, one false-negative, and two false-positive sample status. The PPA, NPA, and OPA with the corresponding two-sided 95% confidence intervals (CI) are summarized in Table 3 and Table 4.

Table 3. Analysis of agreement of overall mutation status per sample: therascreen EGFR Plus RGQ PCR Kit and reference method comparison for plasma samples

    Lower limit 95% CI Upper limit 95% CI
OPA 91.51% 84.49% 96.04%
PPA (sensitivity) 87.27% 75.52% 94.73%
NPA (specificity) 96.08% 86.54% 99.52%

Table 4. Analysis of agreement of overall mutation status per sample after discordant investigation for plasma samples

    Lower limit 95% CI Upper limit 95% CI
OPA 97.17% 91.95% 99.41%
PPA (sensitivity) 97.96% 89.15% 99.95%
NPA (specificity) 96.49% 89.89% 99.57%

Principle

The therascreen EGFR Plus RGQ PCR Kit comprises four triplex assays, each containing all the oligonucleotides (primers, probes and oligonucleotide clamp, where applicable) necessary to amplify and detect the targeted EGFR mutations. In each assay, two target mutations and one internal amplification control are assayed to determine the presence or absence of mutant DNA and to identify reaction failures due to suboptimal DNA input, or the presence of inhibitory substances in the sample matrix.

Allele-specific technology allows accurate and highly reproducible detection of mutations. DNA is selectively amplified using ARMS primers, probes and PCR clamps, with sensitive signal detection using the Rotor-Gene Q MDx 5plex HRM instrument. Result reporting is fully automated using Rotor-Gene AssayManager v2.1 software. If both the positive and no template controls are valid and the sample internal controls are valid, the EGFR mutation status will be displayed in the software.

Procedure

Flexibility is key to the therascreen EGFR Plus RGQ PCR System – with a choice of sample types and DNA extraction methods before real-time PCR.

DNA extraction from FFPE tissue or plasma samples can be performed manually or automated on the QIAsymphony SP, for walkaway sample processing. DNA is extracted from FFPE tissue samples manually using the QIAamp DSP DNA FFPE Tissue Kit, while automated processing uses the QIAsymphony DSP DNA Mini Kit. Plasma samples are processed manually using the QIAamp DSP Circulating Nucleic Acid Kit, while the QIAsymphony DSP Circulating DNA Kit offers automated DNA purification.

Following DNA extraction and quantification, sensitive real-time PCR is performed on the Rotor-Gene Q MDx 5plex HRM instrument. Automated data analysis displays qualitative results in Rotor-Gene AssayManager software, informing the system operator if one or more of the 42 mutations detected by the kit are present. The assay can be completed in ~8 hours, providing next-day results and informing earlier treatment decisions.

Applications

The therascreen EGFR Plus RGQ PCR Kit enables qualitative detection of 42 mutations in the EGFR gene for in vitro diagnostic use. It is an IVD assay to support treatment decisions when selecting NSCLC patients for therapy with an EGFR tyrosine kinase inhibitor (TKI).

Supporting data and figures

Resources

Instrument User Manuals (7)
For installation and setup of Rotor-Gene AssayManager
For use of Rotor-Gene AssayManager in User Defined Test (UDT) mode in combination with the UDT Basic Plug-in
For creation of Rotor-Gene AssayManager assay profiles (only available for the User Defined Test [UDT] mode)
Rotor-Gene Q MDx User Manual, revision R3, for Rotor-Gene Q MDx instruments using Rotor-Gene Q Software version 2.3.4
For use with Rotor-Gene Q Software version 2.3.4
For general use of Rotor-Gene AssayManager
Safety Data Sheets (7)
Download Safety Data Sheets for QIAGEN product components.
Download Safety Data Sheets for QIAGEN product components.
Download Safety Data Sheets for QIAGEN product components.
Brochures & Guides (1)
High-quality, nucleic acid purification for successful PCR and NGS experiments.
Kit Handbooks (3)
Improvements made to your QIAsymphony DSP Circulating DNA Kit
For verification of thermal accuracy of Rotor-Gene real-time cyclers
Protocol Files (1)
Operating Software (1)
For use on the Rotor-Gene Q. Rotor-Gene Q software 2.3.5 is compatible with Windows 7 and Windows 10 operating systems
Labware Documents (1)
For use with software version 5.0 or higher
Technical Information (2)
Certificates of Analysis (1)