QIAxcel RNA Accessories

配合QIAxcel用于RNA分析

S_1084_5_GEN_V2

✓ 全天候自动处理在线订单

✓ 博学专业的产品和技术支持

✓ 快速可靠的(再)订购

QX RNA Denaturation Buffer (2 x 1.4 ml)

目录编号 / ID.   929607

2 x 1.4 ml QX RNA Denaturation Buffer
Accessories
Buffer
Marker
Plasticware
Instrument Accessories
缓冲液类型
QX RNA
QX Buffer
用于
Denaturation
Dilution
QIAxcel RNA Accessories 旨在用于分子生物学应用。这些产品不能用于疾病诊断、预防和治疗。

✓ 全天候自动处理在线订单

✓ 博学专业的产品和技术支持

✓ 快速可靠的(再)订购

特点

  • Accessories for automated capillary electrophoresis of RNA samples
  • For use with QIAxcel RNA Kits and a QIAxcel capillary electrophoresis instrument
  • Safety, convenience and a streamlined workflow

产品详情

QIAxcel附件配合QIAxcel仪器使用,用于全自动进行灵敏、高分辨率的毛细管电泳,每次可处理多达96个样本。

常见问答

When should an intensity calibration be re-performed on the QIAxcel System?

Data for each cartridge intensity calibration on the QIAxcel System is stored in a single file named 'calibration.log'. This file is saved in the BioCalculator root directory C:\Program Files\BioCalculator.

If, for any reason, a different computer is used than the one containing the calibration.log file, the file should be transferred to the new computer. Otherwise, recalibration of the cartridge is required.

Likewise, if the QIAxcel Gel cartridge is used on a different instrument than the one it was calibrated on, another intensity calibration should be performed.

FAQ-1825
What can be done if lower and upper alignment markers used with the QIAxcel System or QIAxcel Advanced are not aligned correctly?

If lower and upper alignment markers used with the QIAxcel System or QIAxcel Advanced are not aligned correctly, open the channel of the sample that was incorrectly aligned. Check whether any unexpected extra peaks have been detected or if the alignment marker peaks do have any “shoulders” that are detected as separate peaks. Delete any extra peaks if necessary and re-analyze your data.

FAQ-1834
How can a clogged capillary channel be identified when running samples on the QIAxcel System or QIAxcel Advanced?

If one of the capillaries of the gel cartridge for the QIAxcel System or QIAxcel Advanced is clogged, the corresponding lane in the gel-view image is typically black.

 

FAQ-1836
Can customers use their own DNA size marker with the QIAxcel System and if yes, what concentration should be used?

The concentration of the DNA marker, pUC18/HaeIII, provided in the QIAxcel DNA Kits is 5ug/ml in 1XTE buffer solution. Customers can run their own DNA size marker sample with different concentration levels using the provided Methods (i.e. M500, H500, L500 and so on).

The recommended method for all basic applications with the QIAxcel System is the M500 method. The BioCalculator software also allows customers to choose their own Sample Injection Time in the Sequence table. The M500 method provides 5KV for 20 seconds for the electro-kinetic injection of lower concentrated DNA samples. If the Injection Time is doubled or tripled (40 or 60 sec), you can expect double or triple intensities (also peak heights).

FAQ-1841