How can I prevent running out of M48 Kit reagents before the full number of samples are processed?
We recommend to process a minimum of 24 samples per run on the BioRobot M48 workstation. While the software will allow as few as 6 samples to be run at a time, small amounts of buffer do remain at the bottom of the reagent trays after each run. This may lead to reagent shortage, unless the remaining buffers are being reused.