The EZ1 Advanced XL sets new standards in data management. Barcode reading enables complete tracking of samples and reagents throughout the entire purification process. Notes and assay identification information can also be easily entered. For increased process control and reliability, reagent identification and lot numbers are logged together with the expiration date, and a warning is issued if the expiration date has been exceeded. The run report, automatically generated at the end of each protocol, contains all relevant instrument information, including maintenance procedures and decontamination data. The instruments generate a pdf and a csv file after each run. Data can be easily transferred to a printer or computer within the network and the report file can be subsequently processed by a LIMS (Laboratory Information Management System).
Proven EZ1 Kits, in combination with EZ1 Advanced XL Cards, ensure high-performance results in purification procedures. The unrivaled ease of use of the EZ1 Advanced XL enables fast and reproducible purification, without the need for specialized training or previous experience with laboratory automation. The extensive application range, including DNA purification from blood, tissues, and human identity and forensic samples, RNA purification from cells and tissues, and purification of viral nucleic acids from a range of sample types, delivers the flexibility that is essential for laboratories.
The EZ1 Advanced XL has been designed to ensure high process safety, high performance, and user convenience. A sensor ensures that the instrument door remains closed during sample preparation, reducing contact of laboratory personnel with potentially infectious samples and protecting sample integrity. At the end of each run, an intelligent-design UV lamp decontaminates the inner surface of the instrument. UV decontamination is highly effective in eliminating Gram-positive and Gram-negative bacteria (see tables). If required, the instrument can be placed in a safety cabinet, providing increased user safety.
Irradiation (min) | Run | 1 | 2 | 3 | 4 | 5 | 6 |
---|---|---|---|---|---|---|---|
30 | 1 | 0 | 0 | 0 | 0 | 0 | 0 |
30 | 2 | 0 | 0 | 0 | 1 | 0 | 0 |
30 | 3 | 10 | 0 | 0 | 0 | 0 | 3 |
30 | Control | 311 | 352 | 367 | 367 | 351 | 396 |
60 | 1 | 0 | 0 | 0 | 0 | 0 | 0 |
60 | 2 | 0 | 0 | 0 | 0 | 0 | 0 |
60 | 3 | 0 | 0 | 0 | 0 | 0 | 0 |
60 | Control | 348 | 307 | 326 | 326 | 396 | 385 |
Six replicates (20 μl each) of Escherichia coli cultures (17,750 CFU/ml) were irradiated using the integrated UV lamp of the EZ1 Advanced, followed by growth on plates in appropriate culture medium. Three independent irradiation runs were performed. Irradiation using the UV lamp was highly effective in eliminating bacteria.
Irradiation (min) | Run | 1 | 2 | 3 | 4 | 5 | 6 |
---|---|---|---|---|---|---|---|
30 | 1 | 12 | 10 | 17 | 2 | 6 | 4 |
30 | 2 | 0 | 0 | 0 | 0 | 0 | 0 |
30 | 3 | 10 | 0 | 2 | 11 | 0 | 3 |
30 | Control | 248 | 256 | 396 | 314 | 321 | 267 |
60 | 1 | 0 | 0 | 0 | 0 | 0 | 0 |
60 | 2 | 0 | 0 | 0 | 0 | 0 | 0 |
60 | 3 | 10 | 0 | 0 | 0 | 0 | 0 |
60 | Control | 264 | 305 | 368 | 292 | 347 | 388 |
Six replicates (20 μl each) of Staphylococcus haemolyticus cultures (15,700 CFU/ml) were irradiated using the integrated UV lamp of the EZ1 Advanced, followed by growth on plates in appropriate culture medium. Three independent irradiation runs were performed. Irradiation using the UV lamp was highly effective in eliminating bacteria.
Automated sample preparation enables efficient, standardized purification of nucleic acids. Manual sample handling is minimized, ensuring safety to the user and reliable processing of samples.
The EZ1 Advanced XL instrument uses proven EZ1 Kits for purification of DNA, RNA, and viral nucleic acids from a wide range of starting materials. Purified nucleic acids are ready for immediate use in downstream applications. The combination of an easy-to-use instrument, with error-free protocol selection and worktable setup, and prefilled, sealed reagent cartridges makes nucleic acid purification easy.
The self-contained EZ1 Advanced XL instrument ensures optimal ease of use and walkaway automation. A wide range of protocols is available for purification of DNA, RNA, or viral nucleic acids. Up to 14 samples can be processed per run in as little as 20 minutes. The instrument is operated via a keypad, and instrument information is displayed on the VFD (vacuum fluorescent display).
All processing steps are performed by the instruments — from piercing reagent cartridges to elution of pure nucleic acids. A separate computer is not required for operation, and EZ1 Kits include all reagents and accessories required to process your samples. EZ1 Kits provide prefilled, foil-sealed reagent cartridges that remain sealed until the instrument door is closed and the protocol run started, reducing the risk of contamination during setup.
If desired, up to 4 EZ1 Advanced XL instruments can be connected to an external computer, expanding throughput up to 56 samples per run. This enables you to easily adapt your laboratory setup to changing sample preparation requirements.
The EZ1 Advanced XL uses proven EZ1 Kits with preprogrammed protocols on EZ1 Advanced XL Cards. EZ1 Kits are available for a broad range of applications including work in human identity and forensics, biomedical research, and gene expression analysis (see table).
Card | Kit | Sample |
---|---|---|
EZ1 Advanced XL DNA Blood Card | EZ1 DNA Blood 200 µl Kit | Blood and blood-derived samples |
EZ1 Advanced XL DNA Blood Card | EZ1 DNA Blood 350 µl Kit | Blood and blood-derived samples |
EZ1 Advanced XL DNA Dried Blood Card | EZ1 DNA Tissue Kit | Dried blood |
EZ1 Advanced XL DNA Buffy Coat Card | EZ1 DNA Blood 350 µl Kit | Buffy coat |
EZ1 Advanced XL DNA Tissue Card | EZ1 DNA Tissue Kit | Tissues |
EZ1 Advanced XL DNA Paraffin Section Card | EZ1 DNA Tissue Kit | Paraffin-embedded tissues |
EZ1 Advanced XL DNA Buccal Swab Card | EZ1 DNA Tissue Kit | Buccal swabs |
EZ1 Advanced XL DNA Investigator Card | EZ1 DNA Investigator Kit | Forensic and human identity samples |
EZ1 Advanced XL DNA Bacteria Card | EZ1 DNA Tissue Kit | Human samples, swabs, biopsies, bacterial cultures |
EZ1 Advanced XL Virus Card v2.0 | EZ1 Virus Mini Kit v2.0 | Viral DNA and RNA and bacterial DNA from serum, plasma, CSF, urine, whole blood, stool, transport media, respiratory samples, and dried swab samples |
EZ1 Advanced XL RNA Card | EZ1 RNA Universal Tissue Kit | Any type of human or animal tissue |
EZ1 Advanced XL DSP DNA Blood Card* | EZ1 DSP DNA Blood Kit* | Human whole blood |
EZ1 Advanced XL DSP Virus Card* | EZ1 DSP Virus Kit* | Viral DNA and RNA and bacterial DNA from serum, plasma, CSF, urine, whole blood, stool, transport media, respiratory samples, and dried swab samples |
Nucleic acids purified using the EZ1 Advanced XL are ready for use in a wide range of downstream applications, including:
Have your throughput demands increased? Do you want to expand your application range or further streamline workflows? Do you need to manage more complex samples? Do you need improved connectivity?
If the answer to any of these questions is yes, then QIAGEN’s trade-in/trade-up program is just right for you!
QIAGEN makes it easy to keep up to speed with the latest automation and sample processing technologies. Simply exchange an old instrument for a new one. Improve your efficiency by trading in a competitor instrument or trading up with an older QIAGEN instrument.
1. Insert the EZ1 Advanced XL Card with the desired protocol into the card slot and start the instrument. 2. Place samples and prefilled, sealed reagent cartridges into the instrument. Samples are processed automatically and purified nucleic acids are transferred to elution tubes. 3. Remove pure nucleic acids in 1.5 ml microcentrifuge tubes at a concentration suited to the downstream application.
Features | Specifications |
---|---|
Features | Touch LCD user interface, IC card with protocols/software, barcode reader, UV lamp |
Pipetting range | 50-1000 µl |
Kits designed for this instrument | QIAGEN EZ1 Kits |
Technology | Magnetic-particle technology |
Instrument dimensions | 510 x 510 x 570 mm (w x d x h) |
Technical data of the instrument | Weight 48 kg, 100–240 V AC, 50–60 Hz |
Protocols/main application on this instrument | Purification of DNA, mRNA, total RNA, and viral RNA and DNA |
Samples per run (throughput) | 1–14 samples per run |
Make sure a correct protocol card is inserted for the corresponding instrument. Protocol cards for BioRobot EZ1, EZ1 Advanced, and EZ1 Advanced XL cannot be interchanged. If you use an incorrect protocol card, you can get error 24.
If you are using a correct protocol card, please contact QIAGEN Technical Service.
The lifetime of the UV lamps is 1200 cycles on an EZ1 Advanced.
The lifetime of the UV lamps is 1500 cycles on an EZ1 Advanced XL.
If the issue persists, please contact QIAGEN Technical Service.
Do not remove protocol cards from the instrument while the instrument is switched on.
If the error 29 occurred on a BioRobot EZ1, you will need to clear the protocol memory and perform teaching. For instructions and teaching data sheet, contact QIAGEN Technical Service.
If the error 29 occurs on an EZ1 Advanced or an EZ1 Advanced XL, restart the instrument. The error should clear. If the error persists, contact QIAGEN Technical Service to obtain teaching instructions and teaching data sheet for your instrument.
Check that the recommended silicone grease (cat. no. 9239334) is used to lubricate the tip adapters on the plunger unit.
Make sure the tip adapters do not have excess grease. Wipe off any excess. Turn off the instrument and perform a test run to see if the error persists.
If correct grease was used and excess grease is not an issue, please contact QIAGEN Technical Service to obtain teaching instructions and teaching data sheet for your instrument.
Instrument seems to have lost teaching. Please contact QIAGEN Technical Service to obtain teaching instructions and teaching data sheet for your instrument.
If issue persists, a service visit may be needed.
Part number for transport lock for EZ1 Advanced XL is 9019723. You can place an order by calling your local QIAGEN customer care.
Instrument may have lost teaching. Please contact QIAGEN Technical Service to obtain teaching instructions and teaching data sheet for your instrument.
If issue persists, a service visit may be needed.
Instrument may have lost teaching. Please contact QIAGEN Technical Service to obtain teaching instructions and teaching data sheet for your instrument.
If issue persists, a service visit may be needed.
Check the workstation for any obstructions or any incorrect placement of the consumables (samples tubes, elution tubes, reagent cartridge, or tips).
If the issue persists, please contact QIAGEN Technical Service to obtain teaching instructions and teaching data sheet for your instrument.
UV lamp needs to be replaced. Please contact QIAGEN Technical Service to schedule service visit.
Instrument may have lost teaching. Please contact QIAGEN Technical Service to obtain teaching instructions and teaching data sheet for your instrument. If teaching does not resolve the issue, a service visit may be needed.
The catalog number for Silicone grease for O-rings of Geno M6, M48, and EZ1 BioRobots is # 9239334. You can place an order by calling your local QIAGEN customer care.